Counting Modes for the Xspress 3
Internal Triggers
- Trigger mode (1): Internal
Utilises the internal time frame generator, therefore frame time is set by PV FrameTime.
Hardware Triggers
- Trigger mode (3): TTL Veto Only
A TTL signal can be sent to the xspress 3 system via the TTL IN 1 connection. When the signal goes high the frame will start and when the signal goes low the signal will stop.
Software Triggers
- Trigger mode (2): Software
Frames can be controlled with the PV ‘SoftTrigger’. Setting 1 to start the frame and 0 to stop the frame and ready the system for the next frame.
- Trigger mode (7): Software + Internal
Frames can be controlled with the PV ‘SoftTrigger’. Setting 1 to start the frame and 0 to ready the system for the next frame. In this mode the frame time is configured by the PV “AcquireTime”. Therefore every frame will be of the same length